21

Working procedures

2.5.1 Procedure for measuring beer colour

1. Prepare samples as described in the working procedure.

2. Select

STORED PROGRAMS

in the “Main Menu”. Select test

number

806

.

3. Insert blank value cuvette (distilled water) into the cell

compartment. Close the cell compartment. Touch

ZERO

.

4. Insert the sample cuvette containing the prepared sample into

the cell compartment and close the cell compartment. Touch

READ

. The result is displayed.

Note: Analysis of additional samples: Repeat working procedure from
point 4.

2.5.1.1 Executing the test with the sipper module

Information about the installation, module configurations and
sample introduction of the sipper module can be found in the user
manual of the DR 5000 Spectrophotometer (15.3.3. Installation
Sipper Module, page 133).