1 executing the test with the sipper module – Hach-Lange DR 5000 brewerie analysis, supplementary software LZV570 User Manual

Page 34

Advertising
background image

34

Working procedures

2.8.1 Procedure for measuring photometric iodine sample

1. Prepare samples and iodine blank solution as described in the

working procedure.

2. Select

STORED PROGRAMS

in the “Main Menu”. Select test

number

810

.

3. Insert cuvette containing phosphate buffer into the cell

compartment. Close the cell compartment. Touch

ZERO

.

4. Insert blank value cuvette containing the prepared iodine blank

solution (see sample preparation) into the cell compartment.
Close the cell compartment. Touch

READ

. Display: E1.

5. Insert the sample cuvette containing the prepared sample into

the cell compartment. Close the cell compartment. Touch

READ

. Display: E2.

6. Introduce 0.5 ml 0.02N iodine solution into sample cuvette. Mix

the contents immediately with plastic spatula, and after 30 sec
insert the cuvette into the cell compartment, close the cover
and touch

READ

. The result is displayed.

Note: Analysis of additional samples: Repeat working procedure from
point 4. The prepared iodine blank solution can be used for the complete
series of measurements.

2.8.1.1 Executing the test with the sipper module

Information about the installation, module configurations and
sample introduction of the sipper module can be found in the user
manual of the DR 5000 Spectrophotometer (15.3.3. Installation
Sipper Module, page 133).