Working procedures

2.6.2 Procedure for measuring free amino nitrogen (FAN) in light beer

1. Prepare three blank value solutions, three standard solutions

and three samples as described in the working procedure.

2. Select

STORED PROGRAMS

in the“Main Menu”. Select test

number

808

.

3. Insert zero value solution (distilled water) into the cell

compartment, close the cover and touch

ZERO

.

4. Insert blank value cuvette (see sample preparation) into the cell

compartment. Close the cell compartment. Touch

READ

.

Display: E1.

Note: Repeat the procedure with blank value cuvettes 2 and 3.
Display: E2 and E3.

5. Insert standard cuvette (see sample preparation) into the cell

compartment. Close the cell compartment. Touch

READ

.

Display: E4.

Note: Repeat the procedure with standard cuvettes 2 and 3. Display:
E5 and E6.

6. Insert the sample cuvette with the first prepared sample into the

cell compartment. Close the cell compartment. Touch

READ

.

Display: E7.

Note: Repeat the procedure with sample cuvettes 2 and 3. Display: E8
and then, after the final measurement, the result.

7. The FAN result is displayed in mg/l.

Note: Analysis of additional samples: Repeat working procedure from
point 6.

2.6.2.1 Executing the test with the sipper module

Information about the installation, module configurations and
sample introduction of the sipper module can be found in the user
manual of the DR 5000 Spectrophotometer (15.3.3. Installation
Sipper Module, page 133).