1 procedure for measuring iron, Working procedures – Hach-Lange DR 5000 brewerie analysis, supplementary software LZV570 User Manual

Page 46

Advertising
background image

46

Working procedures

4. Prepare a blank value solution in exactly the same way, but

without adding the Ferrozin reagent. Prepare a blank value
solution for each beer.

5. Measure the absorbance of the solution in a 40 mm rectangular

cuvette at 560 nm against the corresponding blank value
solution.

Results
The results are expressed in mg/l to three significant places

Accuracy
r = 0.008

Required value
< 0.200 mg/l

Literature
MEBAK Brautechnische Analysenmethoden
4th Edition 2002 Volume II, pp 149ff

2.12.1 Procedure for measuring iron

1. Prepare samples and blank value solution as described in the

working procedure.

2. Select

STORED PROGRAMS

in the “Main Menu”. Select test

number

818

.

3. Insert blank value cuvette (see sample preparation) into the cell

compartment. Close the cell compartment. Touch

ZERO

.

4. Insert the sample cuvette containing the prepared sample into

the cell compartment. Close the cell compartment. Touch

READ

. The result is displayed.

Note: Analysis of additional samples: Repeat working procedure from
point 3.