1 executing the test with the sipper module, Working procedures – Hach-Lange DR 5000 brewerie analysis, supplementary software LZV570 User Manual

Page 37

Advertising
background image

37

Working procedures

Note: Analysis of additional samples: Repeat working procedure from
point 4.

2.9.1.1 Executing the test with the sipper module

Information about the installation, module configurations and
sample introduction of the sipper module can be found in the user
manual of the DR 5000 Spectrophotometer (15.3.3. Installation
Sipper Module, page 133).

2.9.2 Procedure for measuring thiobarbituric acid number in congress wort

1. Prepare samples and blank value solution as described in the

working procedure.

2. Select

STORED PROGRAM

s in the “Main Menu”. Select test

number

812

.

3. Insert zero value cuvette containing distilled water into the cell

compartment. Close the cell compartment. Touch

ZERO

.

4. Insert blank value cuvette (see sample preparation) into the cell

compartment. Close the cell compartment. Touch

READ

.

Display: E1.

5. Insert the sample cuvette containing the prepared sample into

the cell compartment. Close the cell compartment. Touch

READ

. The result is displayed.

Note: Analysis of additional samples: Repeat working procedure from
point 4.