1 executing the test with the sipper module, Working procedures – Hach-Lange DR 5000 brewerie analysis, supplementary software LZV570 User Manual

Page 27

Advertising
background image

27

Working procedures

2.6.3.1 Executing the test with the sipper module

Information about the installation, module configurations and
sample introduction of the sipper module can be found in the user
manual of the DR 5000 Spectrophotometer (15.3.3. Installation
Sipper Module, page 133).

2.6.4 Procedure for measuring free amino nitrogen (FAN) in dark beers

1. Prepare three blank value solutions, three standard solutions

and three samples as described in the working procedure.

2. Select

STORED PROGRAMS

in the “Main Menu”. Select test

number

817

.

3. Insert zero value solution (distilled water) into the cell

compartment, close the cover and touch

ZERO

.

4. Insert blank value cuvette (see sample preparation) into the cell

compartment. Close the cell compartment. Touch

READ

.

Display: E1.

Note: Repeat the procedure with blank value cuvettes 2 and 3.
Display: E2 and E3.

5. Insert standard cuvette (see sample preparation) into the cell

compartment. Close the cell compartment. Touch

READ

.

Display: E4.

Note: Repeat the procedure with standard cuvettes 2 and 3. Display:
E5 and E6.

6. Insert correction cuvette (see sample preparation) into the cell

compartment. Close the cell compartment. Touch

READ

.

Display: E7.

Note: Repeat the procedure with correction cuvettes 2 and 3. Display:
E8 and E9.

7. Insert the sample cuvette with the first prepared sample into the

cell compartment. Close the cell compartment. Touch

READ

.

Display: E10.

Note: Repeat the procedure with sample cuvettes 2 and 3. Display:
E11 and then, after the final measurement, the result.

Advertising