4 sensor membrane specifications, 1 hydrogen sensors, Technical specifications – Hach-Lange ORBISPHERE 31xxx Installation and Maintenance User Manual

Page 12

Advertising
background image

10

Technical Specifications

2.3 EC sensors and parts used on ORBISPHERE configured systems

2.4 Sensor membrane specifications

2.4.1 Hydrogen sensors

Table 2 ORBISPHERE configured systems

Configured System

Sensor

Membrane

Protection

Cap

Protection

Cap Kit

Maintenance

Kit

3625 Package

analyzer

31 110.02

2956A

29111 (as used

on TC sensors)

28002 Tefzel

washers only

32703

3624 ProBrix

31 110.02

2952A

29104.0

29054

32702

29981 Pharmapack

31 110A.02

2956A plus one

Goretex

®

membrane ref.

32918

None

Not applicable

32703 plus

32918

(10 piece box)

Table 3 Membrane specifications - Hydrogen sensors

Membrane Model

2956A

2952A

2995A

29015A

Recommended
applications

Trace

measurement

Low

concentration

Average

concentration

High

concentration

Material PFA

Tefzel

®

Tedlar

®

Saran

Thickness [µm]

25

25

12.5

23

Calibration gas

1% pure H

2

10% pure H

2

100%

pure

H

2

100% pure H

2

Dissolved measurement
range

0 ppb to

75 ppb

0 ppb to
300 ppb

0 ppb to

3200 ppb

0 ppb to

32 ppm

Gaseous measurement
range

0 Pa to

5 kPa

0 Pa to
20 kPa

0 Pa to

200 kPa

0 kPa to

2000 kPa

Accuracy

The greater of ±1%

of reading

or ± 0.03 ppb,

or ± 1 Pa

The greater of ±1%

of reading

or ± 0.09 ppb,

or ± 6 Pa

The greater of ±1%

of reading

or ± 1 ppb,

or ± 50 Pa

The greater of ±1%

of reading

or ± 10 ppb,

or ± 1 kPa

Integrated radiation dose
limit

2 x 10

4

10

8

10

8

N/A

Expected current in air @
1 bar 25°C [µA]

N/A

Expected current in pure
gas [µA]

150

50

5

0.5

Temp. compensation
range

0 to 50°C

0 to 50°C

10 to 45°C

10 to 45 °C

Temp. measuring range

–5 to 100° C

Response time

1

2 sec.

5 sec.

6 sec.

50 sec.

Recommended min. liquid
flow rate

2

[mL/min]

50 to 220

40 to 200

20 to 70

20 to 40

Recommended min. linear
flow rate

2

[cm/sec]

200

150

50

30

Recommended gaseous
flow rate [L/min]

0.005 to 3

Advertising