4 ozone sensors, Technical specifications 2.4.4 ozone sensors – Hach-Lange ORBISPHERE 31xxx Installation and Maintenance User Manual

Page 15

Advertising
background image

13

Technical Specifications

2.4.4 Ozone sensors

Table 6 Membrane specifications - Ozone sensors

Membrane Model

2956A

29552A

Recommended
applications

Trace

measurement

High concentration

(> 1 mg/l)

Material PFA

PTFE

Thickness [µm]

25

50

Calibration gas

Span gas

or air

Dissolved measurement
range

0 ppb to

50 ppm

0 ppb to

200 ppm

Gaseous measurement
range

0 Pa to
10 kPa

0 Pa to

40 kPa

Accuracy

The greater of ±1%

of reading

(± 5% for sensors

calibrated in air)

or ± 5 ppb,

or ±1 Pa

The greater of ±1%

of reading

(± 5% for sensors

calibrated in air)

or ± 20 ppb,

or ± 4 Pa

Integrated radiation dose
limit

2 x 10

4

N/A

Expected current in air @
1 bar 25°C [µA]

26.4

6.5

Expected current in pure
gas [µA]

105

31.4

Temp. compensation
range

–5 to 45° C

Temp. measuring range

–5 to 100° C

Response time

1

1

Response time at 25 °C for a 90% signal change

30 sec.

6 min.

Recommended min. liquid
flow rate

2

[mL/min]

2

Liquid flow through an ORBISPHERE 32001 flow chamber, with

protection cap and no grille

350

3

3

These flow rates take into account the decomposition of ozone in

the tubing between the line and the flow chamber (theoretical flow
rates in the absence of decomposition would be 10 times less)

100

3

Recommended min. linear
flow rate

2

[cm/sec]

30

10

Recommended gaseous
flow rate [L/min]

0.01 to 3