Installation

4.3 Sensor insertion

Insert the sensor straight into the flow chamber or socket.

Hand tighten the attaching collar.

Connect the sensor cable.

Check for leaks; replace O-rings if product leaks are visible.

Micro volume flow chambers:
Do not twist the sensor when inserting it into a micro volume flow chamber. This rotation may
twist the membrane holding ring, thus changing the membrane position. This can modify the
membrane measuring conditions, and affect measurement precision.

Note:

Check that the small O-ring at the bottom of the flow chamber is present during removal and

installation of the sensor, as it may stick to the sensor head and fall.

4.4 Sensor removal

Shut off the sample flow and drain the sampling circuit of liquid or gas.

Remove the sensor cable connected at the sensor end.

Hold the sensor body in one hand to avoid rotation and unscrew the collar with the other.

Pull the sensor straight out of the socket or flow chamber.

Check that both O-rings remain in place inside the flow chambers.

Install sensor storage cap and sensor base (to protect the connection).

4.5 Mounting accessories

Note:

Check the spare part list at the end of this manual for ordering information.

4.5.1 External pressure sensor

The system can be fitted with an external pressure sensor. This enables a measurement of
fraction of gas under variable pressure conditions during gas phase measurement.

Figure 6 Sensor insertion

Two models are available, depending on the application:

28117 Pressure sensor 0 - 5 bar absolute

28117C Pressure sensor 0 - 1 bar absolute