Technical Specifications

2.4.2 Oxygen sensors (Table 1)

Table 4 Membrane specifications - Oxygen sensors (1)

Membrane Model

2956A

2958A

29552A

2952A

Recommended
applications

Corrosion control,

De-aerated water

Beverage,

Lab. applications

In line wort,

Air/O

2

injection,

Sewage treatment

Corrosion control,

In line beverage,

De-aerated water

Material PFA

Tefzel

®

PTFE

Tefzel

®

Thickness [µm]

25

12.5

50

25

Calibration gas

Air

Air

Air

Air / Pure O

2

Dissolved measurement
range

0 ppb to

20 ppm

0 ppb to

40 ppm

0 ppb to

80 ppm

0 ppb to

80 ppm

Gaseous measurement
range

0 Pa to

50 kPa

0 Pa to

100 kPa

0 Pa to

200 kPa

0 Pa to

200 kPa

Accuracy

The greater of ±1%

of reading

or ± 0.1 ppb

1

,

or ± 1 ppb

2

,

or ± 0.25 Pa

The greater of ±1%

of reading

or ± 1 ppb,

or ± 2 Pa

The greater of ±1%

of reading

or ± 2 ppb,

or ± 5 Pa

The greater of ±1%

of reading

or ± 2 ppb,

or ± 5 Pa

1

Accuracy is ± 0.1 ppb for 410, 510, 362x, 360x and 3655 instruments

2

Accuracy is ± 1 ppb for 366x and 3650 instruments

Integrated radiation dose
limit [rads]

2 x 10

4

10

8

N/A

10

8

Expected current in air @
1 bar 25°C [µA]

26.4

9.4

6.3

5.4

Expected current in pure
O

2

[µA]

132

47

31.4

27

O

2

consumption in O

2

saturated water at 25°C
[µg/hour]

40

14

9.4

8

Temp. compensation
range

– 5 to 60° C

Temp. measuring range

– 5 to 100° C

Response time

1

7.2 sec.

9.5 sec.

90 sec.

38 sec.

Recommended min. liquid
flow rate

2

[mL/min]

180

120

50

50

Recommended min. linear
flow rate

2

[cm/sec]

200

100

30

30

Recommended gaseous
flow rate [L/min]

0.1 to 3