3 oxygen sensors (table 2) – Hach-Lange ORBISPHERE 31xxx Installation and Maintenance User Manual

Page 14

Advertising
background image

12

Technical Specifications

2.4.3 Oxygen sensors (Table 2)

Table 5 Membrane specifications - Oxygen sensors (2)

Membrane Model

2935A

29521A

2995A

Recommended
applications

Saturated to super

saturated levels

Saturated to super

saturated levels

In line hot wort

(up to 70°C)

Material Halar

®

Tefzel

®

Tedlar

®

Thickness [µm]

25

125

12.5

Calibration gas

Air / Pure O

2

Air / Pure O

2

Pure

O

2

Dissolved measurement
range

0 ppb to

400 ppm

0 ppb to

400 ppm

0 ppb to 2000 ppm

Gaseous measurement
range

0 Pa to

1000 kPa

0 Pa to

1000 kPa

0 Pa to 5000 kPa

Accuracy

The greater of ±1%

of reading

or ± 10 ppb,

or ± 20 Pa

The greater of ±1%

of reading

or ± 10 ppb,

or ± 20 Pa

The greater of ±1%

of reading

or ± 50 ppb,

or ± 100 Pa

Integrated radiation dose
limit [rads]

N/A

10

8

10

8

Expected current in air @
1 bar 25°C [µA]

0.9

0.7

0.2

Expected current in pure
O

2

[µA]

4.7

3.8

0.9

O

2

consumption in O

2

saturated water at 25°C
[µg/hour]

1.4

1.3

0.3

Temp. compensation
range

– 5 to 60° C

Temp. measuring range

– 5 to 100° C

Response time

1

2.5 min.

18 min.

80 sec.

Recommended min. liquid
flow rate

2

[mL/min]

25

25

5

Recommended min. linear
flow rate

2

[cm/sec]

20

60

5

Recommended gaseous
flow rate [L/min]

0.1 to 3