Introduction – Hach-Lange ORBISPHERE 31xxx Installation and Maintenance User Manual

Page 18

Advertising
background image

16

Introduction

The sensor is designed with one center electrode (cathode) and one counter electrode (anode)
immersed in an electrolytic solution. The electrodes and the electrolyte are separated from the
gaseous or liquid sample by a membrane permeable to gas.

Note:

Anode and cathode positions are inverted on the H

2

sensor

Additionally, the sensor's design includes a guard ring electrode which surrounds the sensor's
center electrode. Its role is to reduce the influence of other gases on the center electrode,
improving analysis stability.

The sensor head is covered with a protection cap and, in some applications, a grille to protect
the membrane. Materials used for the components of the sensors differ with the application.

Gas penetrating through the membrane into the cell dissolves in the electrolyte. It undergoes a
reaction at the cathode, causing a measurable electric current to flow. This current is
proportional to the amount of gas entering the cell. The amount of gas entering the cell is
proportional to the partial pressure of this gas in the sample, outside the cell.

The result is shown as gas concentration, which can then be displayed with a choice of several
measuring units, according to instrument setup.

The sensor electronics perform four functions:

Apply constant voltage to the anode

Measure the current flowing through the sensor

Compensate for temperature variation in the gaseous or liquid sample

Convert the cell’s electric current into an analog signal for sensor output

Electrochemical sensor assembly

Protection cap and cable removed

Sensor cell, with membrane and electrolyte

Membrane and its support removed

Figure 4 Electrochemical sensor overview

Advertising