3 ozone sensor, Troubleshooting – Hach-Lange ORBISPHERE 31xxx Installation and Maintenance User Manual

Page 38

Advertising
background image

36

Troubleshooting

6.3 Ozone sensor

When the O

3

sensor has been properly calibrated using the ORBISPHERE measuring

instrument, the sensor has to settle down for up to 24 hours when used in very low O

3

concentration conditions.

Table 10 Troubleshooting - Ozone sensor

Problem

Probable Cause

Possible Solution

Sensor won't calibrate,
even after thorough
servicing.

Repeated calibrations go beyond
"expected limits" of instrument.

MOCA 3600 only:
Select membrane from
"Options/Membrane" menu. Then,
calibrate the sensor.
O

3

sensor cleaned with HNO

3

Instrument internal barometric
pressure sensor needs
calibration.

Calibrate "Barom. Pressure" menu.
Check against a certified barometer. Do
not correct for sea level !

Wet membrane interface.

Wipe dry with a tissue and re-calibrate.

"0000" O

3

levels

displayed.

Wrong reading scale "XXXX"
selected for display unit.

Change reading scale by selecting
"X.XXX, XX.XX or XXX.X" from
"Options/Display Units" menu.

Unexpected/incorrect
dissolved O

3

reading.

High residual current.

If concentration is significantly higher
than low limit, try a sensor service.

Insufficient flow rate.

Regulate flow equivalent to membrane
specified levels.

Length of sample line allows O

3

time to react.

Reduce length of sample tubing.

Doesn't match lab samples.

Take samples at close proximity to
sensor.