Installation

The connection diagram below is a recommended installation that allows for measuring and
calibrating without having to disconnect a line manually. "A" and "B" represent 3-way valves.

For measuring, calibration gas inlets and outlets are shut off. During calibration, the flow is
reversed to drive the remaining sample out. The calibration gas enters at the "sample out" port
and exits at the "sample in" port, as shown.

Table 7 Flow chamber orientation

Sample

Orientation of flow chamber

Gaseous or
liquid media

Vertically, with connections down and
sensor up

Center connection is the inlet

Outer connection is the outlet

Gaseous media,
with occasional
liquid or vapor

Horizontally, to allow for drainage

Center connection (inlet) must be up

Outer connection (outlet) must be
down

Figure 10 Flow chamber connections