1 general operation, 2 starting sample flow – Hach-Lange SURFACE SCATTER 7 sc Turbidimeter User Manual

Page 29

Advertising
background image

27

Section 4

System startup

4.1 General operation

1. Plug the SS7 sc/SS7 sc-HST into the unpowered controller by

aligning the orientation tab on the cable connector with the
channel in the controller connector.

2. Push in and turn the threaded collar to secure the connection.

Tug gently to check the connection.

3. After all plumbing and electrical connections have been

completed and checked, supply power to the system.

4. Ensure the sample unit door is securely latched when power is

applied, since dark readings are measured at this time. If power
is applied while the door is open, cycle the power with the door
closed. The dark readings are measured again one hour after
the power-up.

5. The first time a controller is powered up, a language selection

menu will appear. Select the correct language from the
displayed options.

6. Following language selection and upon power-up, the controller

will search for connected sensors. The display will show the
main measurement screen.

4.2 Starting sample flow

1. Start sample flow through the instrument by opening the

sample supply valve.

2. Allow the turbidimeter to run long enough for the tubing and

body to become completely wetted and the reading on the
display to stabilize. One to two hours or longer may be required
initially for complete stabilization.

3. Allow measurements to become stable through adequate

conditioning before completing instrument settings or
performing calibrations.