Hach-Lange SURFACE SCATTER 7 sc Turbidimeter User Manual

Page 7

Advertising
background image

5

Section 1

Specifications

Specifications are subject to change without notice

Range

0.01–9999.9 nephelometric turbidity units (NTU)

Accuracy

± 5% of reading or ± 0.1 NTU (whichever is greater) from 0.01 to 2000 NTU;
± 10% of reading from 2000 to 9999 NTU

Resolution (displayed)

0.01 NTU

up to 999.99 NTU; 0.1 NTU from 1000.0 to 9999.9 NTU

Repeatability

Better than ± 1.0% of reading or ± 0.04 NTU, whichever is greater for each range.

Response time

Initial response in 45 seconds

Sample flow required

1.0 to 2.0 L/min (0.3 to 0.5 gal/min) (15 to 30 gal/hr)

Sensor storage
temperature

–20 to 80 °C (–4 to 140 °F); 95% relative humidity, non-condensing.

Operating temperature

sc100: 0 to 50 °C (32 to 122 °F) for one SS7 sc on an sc100; 0 to 40 °C (32 to 104 ° F) for one
SS7 sc and another smart sensor that consumes less than 5 watts on a single sc100
(provided by sc100). See

Figure 1 on page 6

.

sc1000: –20 to 55 °C (–4 to 131 °F); 95% relative humidity, non-condensing

Sample temperature range

0 to 50 °C (32–122 °F);
HST model—0 to 70 °C, intermittent 70 to 80 °C. (An approved heat exchanger is available to
reduce sample temperature.)

Operating humidity

5 to 95% non-condensing

Power requirements

12 VDC ±5%, 20 watts maximum (provides by sc100)

Sample inlet fitting

¾-in. NPT female

Overflow drain fitting

1-in. NPT female

Body drain fitting

¾-in. NPT female

AIr purge fitting

¼-in. quick-connect compression fitting; 0–50SCFH airflow of clean instrument air

Signal average (filter) time

No averaging, 6, 30, 60 and 90 seconds, user selectable. Default is 30 seconds.

Sensor dimensions

64.2 x 67.5 x 19.0 cm (25.3 x 26.6 x 7.5 in.)

Sensor cable length

2 m (6.6 ft); Optional 7.62 m (25 ft) extension cable. Maximum cable length is 9.62 m (31.6 ft).

Sensor cable rating

Cable: 105 °C, 300 V, PVC jacket
Wires: 22 AWG, PVC jacket

Mounting options

Wall

Shipping weight

SS7 sc—15.8 kg (34.8 lb);
SS7 sc-HST—18 kg (39.6 lb)

Calibration method

Formazin – user-prepared primary or wet calibration of the instrument

Verification (dry) method

Standardization plates with approximate values of 100 or 1000 NTU. Unique value is assigned
when dry verification is done immediately after calibration and is used with pass/fail criteria for
subsequent verifications.

Recommended cleaning
intervals

Mandatory before calibration
Optional before verification
Mandatory upon verification failure

Languages

English (default), German, French, Spanish, Italian, Swedish, Polish, Korean,
Chinese, Japanese

Installation environment

Indoor

Primary compliance
method

USEPA 180.1; Hach Method 8195; ASTM D 6698; Standard Methods 2130B