2 sensor data logging, 3 sensor diagnostics menu, 4 sensor setup menu – Hach-Lange SURFACE SCATTER 7 sc Turbidimeter User Manual

Page 32: Operation

Advertising
background image

30

Operation

5.2 Sensor data logging

The controller provides two data logs (one for each sensor) and two
event logs (one for each sensor). The data logs store the
measurement data at selected intervals. The event log stores a
variety of events that occur on the devices such as configuration
changes, alarms and warning conditions. The data logs are stored
in a packed binary format and the event logs are stored in a CSV
format. The logs can be downloaded through the digital network
port, service port or the IrDA port. DataCom (Cat. No. 59256-00 or
download from www.hach.com) is needed for downloading logs to a
computer. If the datalogging frequency is set to 15 minute intervals,
the instrument can continue to store data for approximately
six months.

1. From the Main Menu, select SENSOR SETUP and confirm.

2. Highlight the appropriate sensor if more than one sensor is

attached and confirm.

3. Select CONFIGURE and confirm.

4. Select the datalog interval (

5 seconds, 30 seconds, 1 minute,

2 minutes, 5 minutes, 10 minutes, 15 minutes, 30 minutes, 60 minutes
or 4 hours

). Confirm.

5.3 Sensor diagnostics menu

SELECT SENSOR

ERROR LIST

See

section 7.1 on page 47

.

WARNING LIST

See

section 7.2 on page 47

.

5.4 Sensor setup menu

SELECT SENSOR (if more than one sensor is attached)

CALIBRATE

PERFORM CAL

Calibration using 4000 NTU stock solution

VERIFICATION

Perform a verification, set the pass/fail criteria and view the verification history.

0 ELECTRONICS

Zero electronics

CAL HISTORY

View the last 12 entered calibrations. Confirm to move to the next history entry. See

section 5.6

on page 36

for more information.

CONFIGURE

BUBBLE REJECT

Choose Yes or No to enable/disable bubble reject. Default: Yes

SIGNAL AVG

Choose no averaging or specify the amount of time for signal averaging. Available options are:
no averaging, 6 sec., 30 sec., 60 sec. or 90 sec. Default is 30 seconds.

MEAS UNITS

Select the appropriate measurement units to display. Choose from mg/L, NTU, FTU and NO
UNITS. Default: NTU

EDIT NAME

Enter up to a 12-digit name in any combination of symbols and alpha or numeric characters.
Confirm when the entry is complete. The name will be displayed on the status line above the
measurement value on the main display. Default is SS7.

SET RESOLUTION

Set the number of significant digits to display. Default is one significant digit.

DATALOG INTRVL

Choose the amount of time between saving data points to the data log. Default: 15 min.;
Options: 5 seconds, 30 seconds, 1 minute, 2 minutes, 5 minutes, 10 minutes, 15 minutes,
30 minutes, 60 minutes or 4 hours.

Advertising