1 error codes, 2 warnings – Hach-Lange SURFACE SCATTER 7 sc Turbidimeter User Manual

Page 49

Advertising
background image

47

Section 7

Troubleshooting

7.1 Error Codes

Errors are indicated by a flashing measurement value and a
flashing warning icon. Errors are defined in

Table 5

.

1. From the Main Menu, select SENSOR DIAG and confirm.

2. If multiple sensors are attached to the controller, choose

SELECT SENSOR>SS7 SETUP and confirm.

3. Select ERROR LIST and confirm. All active errors will display.

7.2 Warnings

Warnings are indicated by a flashing measurement value and a
flashing warning icon. Warnings are defined in

Table 6

.

1. From the Main Menu, select SENSOR DIAG and confirm.

2. If multiple sensors are attached to the controller, choose

SELECT SENSOR>SS7 SETUP and confirm.

3. Select WARNING LIST and confirm. All active warnings

will display.

Table 5 Error codes

DIsplayed error

Definition

ADC FAIL

The ADC has failed. Try cycling power. If cycling power does not work, replace the detector
assembly (Cat. No. 71221-00).

LAMP FAIL

The light source has failed. See

section 6.4.1 on page 40

for lamp replacement instructions.

FLASH FAIL

Datalog and event log will not work.

Table 6 Warning Codes

Warning

Number

Displayed Warning

Definition/Resolution

1

DARK WARNING

Dark reading detects too much light. Close the SS7 sc enclosure and perform
ZERO ELECTRONICS (under the CALIBRATION menu).

2

TEMP WARNING

Sensor head internal temperature is higher than specified. Contact the
Technical Support Department. (> 70 °C)

3

DATA LOG FULL

Sensor data log is full. No additional data will be logged until sensor log is
downloaded into controller memory.

4

EVENT LOG FULL

Sensor data log is full. No additional data will be logged until sensor log is
downloaded into controller memory.

5

5 VOLT WARN

Monitored voltage is outside the range of 4.5–5.5 V.

6

VIN WARN

Monitored instrument input voltage from sc100 is outside the range of
9.08–14.3 V. Check cables. Make sure only one SS7 sc is connected to sc100
and if any other probe is connected it can only draw 4 watts maximum.

7

LAMP VOLT WARN

Monitored voltage is outside the range of 3.96–4.48 V.

8

LAMP CURR WARN

Monitored current is outside the range of 1.67–2.75 Amps.

9

OUTPUT MODE WARN

Activated when the sensor is not in normal measurement mode (such as when
in calibration or verification mode).

10

AC UPDATE FAIL

The application code update failed.

11

EXT FLASH FAIL

External copy of the application code has failed. Self recovery should occur.