1 scheduled maintenance, 2 removing a sensor from the system, 3 installing a sensor on the system – Hach-Lange SURFACE SCATTER 7 sc Turbidimeter User Manual

Page 41

Advertising
background image

39

Section 6

Maintenance

DANGER
Only qualified personnel should conduct the tasks described
in this section of the manual.

The nature of tasks described in this section of the manual requires
individuals to be technically knowledgeable of the associated
dangers. Burns, shock, eye damage, fire and chemical exposure
may occur if this work is not done by qualified personnel. Always
review appropriate Material Safety Data Sheets (MSDS) before
working with chemicals.

6.1 Scheduled maintenance

Scheduled periodic maintenance requirements of the SS7 sc
Turbidimeter are minimal. Standardization checks and calibration
are the primary requirements. Several other activities should be
performed on a regular basis, but the schedule for these may
depend on the installation and sample.

6.2 Removing a sensor from the system

Prior to physically removing a sensor from the system, record all
user defined settings such as relays, signal averaging, etc. Turn off
power to the sc100 and SS7 sc, then disconnect the sensor at
the controller.

6.3 Installing a sensor on the system

To return the system to normal operation following a software
upgrade or sensor repair, perform the following procedure:

1. Detach all sensors from the sc100 controller.

2. From the Main Menu, press the

DOWN

key to highlight

TEST/MAINT and confirm.

3. Use the

DOWN

key to scroll to SCAN SENSORS and confirm.

4. Remove attached sensors by selecting the corresponding serial

number or select “All”.

5. Power down the sc100 controller, then attach the sensor(s)

to be used.

Note: Clean sensors before installing on the system.

6. Supply power to the sc100 controller. The system will

initialize automatically.

Advertising