1 connectors, 1 selection of diaphragm plate, Installation 3.1.1 connectors – Hach-Lange ULTRATURB sc basic_plus_seawater User Manual
Page 12

12
Installation
3.1.1 Connectors
Important Note: The instrument can be damaged if the connectors and/or the
union nuts are attached too tightly. Only hand-tighten the union nuts for the hose
mounts. If necessary, hold the connectors with a tool (SW 22).
The measuring instrument can be integrated into the measurement and analysis
process using tubing (ID 13 mm) or a fixed connection (PVC system component
pipework, compatible with 1" union nut).
3.1.1.1 Selection of diaphragm plate
With samples that tend to emit gases, bubbles may form in the measuring
chamber. This causes extremely strongly fluctuating measurement values. In the
event of strongly fluctuating measurement values, replace the diaphragm plate in
the drain connector.
Note: Observe the flow quantity and flow direction of the sample provided in the technical
data.
Table 1: Connectors
External thread
Pipe thread in accordance with DIN ISO 228
large
G1A
small
G½A
Table 2: Preselect the diaphragm plate
Flow quantity/pressure
Diaphragm plate
large
large hole
medium
medium hole
small
small hole