Section 5 maintenance, 1 maintenance schedule, 2 cleaning measuring chamber – Hach-Lange ULTRATURB sc basic_plus_seawater User Manual

Page 21

Advertising
background image

21

Section 5

Maintenance

It has been possible to reduce the scope of the maintenance to be performed by
the user to only a few actions. Clearly summarised in a table and described in
detail in the following sections, these actions can be quickly and straightforwardly
performed by qualified personnel.

5.1 Maintenance schedule

5.2 Cleaning measuring chamber

The cleanliness of the measuring chamber is crucial for correct measured results.
This cylindrical optical measuring chamber has a rotating wiper with three wiper
profiles (only on plus and seawater version); these prevent normal soiling of the
optical system as it starts to build up. With very stubborn deposits, manual
cleaning with a suitable cleaning agent (e.g. citric acid) is recommended.

CAUTION
Observe safety regulations and wear safety clothing!

Safety glasses

Gloves

Overalls

1. Open the MAIN MENU.

2. Choose SENSOR SETUP and accept.

3. Choose the related sensor and accept.

4. Choose TEST/MAINT and accept.

5. Choose MAINT. PROC and accept.

6. Choose CLN. MEAS CHAMB and accept.

7. Shut-off the feed and accept TURN OFF SAMPLE INLET.

8. Drain the measuring chamber via the lower feed union and accept DRAIN

MEAS. CHAMBER.

Maintenance task

Interval

Clean measuring chamber:

Dependent on the substances contained

in the water

Replace wiper profile
(only plus and seawater version)

After 1200 cycles

Replace desiccant

Every 2 years

Monitor test equipment (CVM dry
standard)

Every 2 years

(factory test with certification)

Check zero point

Dependent on the substances contained

in the water

Check gradient

At least once a year