5 calibration with standard solution, Operation – Hach-Lange ULTRATURB sc basic_plus_seawater User Manual

Page 18

Advertising
background image

18

Operation

4.5 Calibration with standard solution

1. Open the MAIN MENU.

2. Choose SENSOR SETUP and accept.

3. Choose the related sensor and accept.

4. Choose CALIBRATE and accept.

5. Choose STANDARD and accept.

6. Shut-off the feed and accept TURN OFF SAMPLE INLET.

7. Drain the measuring chamber using the lower feed union. Accept DRAIN

MEAS. CHAMBER.

8. Connect the calibration syringe to the feed union (see Figure) and add the

calibration standard. Accept POUR STD INTO MEAS. CHAMBER.

9. Accept PRESS ENTER WHEN STABLE x.xxx TRBFNU.

10. Enter the concentration of the standard solution. Accept CALIBRATE

(x.xxx TRBFNU).

11. Remove the calibration syringe and accept REMOVE CALIBRATION

SYRINGE.

12. Re-connect the feed and accept CONNECT SAMPLE INLET.

13. Open the feed and accept CAL READY OPEN INLET.

14. Accept OUTPUT ACTIVE.