Operation
4.4.2 SENSOR SETUP
WIPE
Initiates a wiping action
VERIFY
Verification using CVM
module
Verification with dry standard
CALIBRATE
STANDARD
Selection as per
CAL. CONFIG
OUTPUT MODE
Calibration with standard
solution
Procedure during calibration, menu
based
OFFSET
Selection as per
CAL. CONFIG
OUTPUT MODE
Zero point setting
Zero point setting, menu based
CAL. FAKTORS
FAC. STANDARD
0.50 to 2.00
OFFSET
-0.100 to +0.100 TRBFNU
CAL. CONFIG
OUTPUT MODE
HOLD
ACTIVE
TRANSFER
CHOICE
Behaviour of the outputs during
calibration or zero point setting
CAL. INTERVAL
Adjustable from 0 to 365 days
SET CAL DEFLT
Reset to the default calibration
CONFIGURE
EDITED NAME
SET DEFAULTS Instrument
number
Up to 16-character name possible
MEAS UNITS
mg/L, FNU, NTU, TE/F,
EBC
SET DEFAULTS
FNU
CLEAN. INTERVAL
SET DEFAULTS
12 h
10 min, 20 min, 30 min,
2 h, 6 h, 12 h, 10:00
RESPONSE TIME
SET DEFAULTS
15 s
0 to 60 s
LOGGER INTERVAL
SET DEFAULTS
10 min
1-30 min
RESOLUTION
SET DEFAULTS
0.xxx
0.xxx or 0.xxxx (<1 FNU)
SET DEFAULTS
Confirmation prompt
Reset to default configuration for all
menu commands listed above.