Operation

4.4.2 SENSOR SETUP

WIPE

Initiates a wiping action

VERIFY

Verification using CVM
module

Verification with dry standard

CALIBRATE

STANDARD

Selection as per
CAL. CONFIG
OUTPUT MODE

Calibration with standard
solution

Procedure during calibration, menu
based

OFFSET

Selection as per
CAL. CONFIG
OUTPUT MODE

Zero point setting

Zero point setting, menu based

CAL. FAKTORS

FAC. STANDARD

0.50 to 2.00

OFFSET

-0.100 to +0.100 TRBFNU

CAL. CONFIG

OUTPUT MODE

HOLD
ACTIVE
TRANSFER
CHOICE

Behaviour of the outputs during
calibration or zero point setting

CAL. INTERVAL

Adjustable from 0 to 365 days

SET CAL DEFLT

Reset to the default calibration

CONFIGURE

EDITED NAME

SET DEFAULTS Instrument
number

Up to 16-character name possible

MEAS UNITS

mg/L, FNU, NTU, TE/F,
EBC

SET DEFAULTS
FNU

CLEAN. INTERVAL

SET DEFAULTS
12 h

10 min, 20 min, 30 min,
2 h, 6 h, 12 h, 10:00

RESPONSE TIME

SET DEFAULTS
15 s

0 to 60 s

LOGGER INTERVAL

SET DEFAULTS
10 min

1-30 min

RESOLUTION

SET DEFAULTS
0.xxx

0.xxx or 0.xxxx (<1 FNU)

SET DEFAULTS

Confirmation prompt

Reset to default configuration for all
menu commands listed above.