Components

Microprocessor-controlled turbidity bypass sensor ULTRATURB sc for
very low to medium turbidities with comprehensive self-diagnostics

Measuring technique

90° infrared pulse scattered light technique in accordance with DIN EN ISO 7027

Measuring range

0.0001–1000 FNU (TE/F, NTU, FTU) can be programmed as required
(0.0001–250 EBC = 2500 ppm SiO2)

Resolution

0.0001–0.9999 / 1.00–9.99 / 10.0–99.9 / 100–1000 FNU

Precision

±0.008 FNU or ±1 % of the measured value (0–10 FNU)

Reproducibility

±0.003 FNU or ±0.5 % of the measured value (0–2 FNU)

Technique variation coefficient

1 % in accordance with DIN 38402

Response time

1–60 s (can be programmed as required)

Air bubble compensation

Physical-mathematical

Calibration

Permanently set in the factory
(Validation using formazine, StablCal, dry standard CVM)

Sample flow rate

Min. 0.2 L/min, max. 1 L/min, max. 6 bar (at 20 °C (at 68 °F))

Sample temperature

Max. 50 °C (Max. 122 °F)

Salt content of the probe
(only for seawater
)

Tested up to 65 g/L

Ambient temperature

+2 °C to +40 °C (+ 36 °F to +104 °F)

Sample connection

Tubing (ID 13 mm) or fixed connection (PVC system component pipework)

Automatic cleaning of the
measuring chamber
(only for plus
and seawater)

Automatic wiper cleaning, time controlled and as required

Materials

Measuring window: quartz
Measuring chamber: Noryl GFN2
Wiper axle: stainless steel 1.4571
Wiperarm (seawater): titanium alloy
Wiper profile: silicone

Housing

Enclosure rating: IP 65
Plastic housing ASA

Inspection interval

Once a year, service contract on request with warranty extension of five years;
please request a quote

Dimensions

(W × H × D) 250 Ч 240 Ч 110 mm

Mass

Approx. 1.5 kg

User maintenance

basic

: 2 h / month, typical

plus

and seawater: 0.5 h / month, typical