Connect the tubing to the sampler as shown in, Figure 6 – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS Basic User Manual User Manual

Page 12

Advertising
background image

Note: Use connection kit 2186 if Teflon-lined tubing is used.

2. Install the intake tube and strainer in the main stream of the sample source where the water is

turbulent and well-mixed. Refer to

Figure 7

.

• Make the intake tube as short as possible.
• Keep the intake tube at a maximum vertical slope so that the tube drains completely between

samples.

Note: If a vertical slope is not possible or if the tube is pressurized, disable the liquid sensor. Calibrate the
sample volume manually.

• Make sure that the intake tube is not pinched.

Figure 6 Intake tube installation

Figure 7 Site installation

1 Strainer

2 Vertical lift

3 Mounting surface