Connect the tubing to the sampler as shown in, Figure 6 – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS Basic User Manual User Manual
Page 12
Advertising
Note: Use connection kit 2186 if Teflon-lined tubing is used.
2. Install the intake tube and strainer in the main stream of the sample source where the water is
turbulent and well-mixed. Refer to
.
• Make the intake tube as short as possible.
• Keep the intake tube at a maximum vertical slope so that the tube drains completely between
samples.
Note: If a vertical slope is not possible or if the tube is pressurized, disable the liquid sensor. Calibrate the
sample volume manually.
• Make sure that the intake tube is not pinched.
Figure 6 Intake tube installation
Figure 7 Site installation
1 Strainer
2 Vertical lift
3 Mounting surface
12 English