Time based collection, Use variable volume, Figure 12 – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS Basic User Manual User Manual
Page 18
Option
Description
Advanced sampling
Goes to the advanced sampling menu. If necessary, configure the options in the
Advanced Sampling menu.
Run program
The creation of the basic sampling program finishes. The sampler prompts the user to
start or cancel the program.
Figure 12 Two samples per bottle (left) versus two bottles per sample (right)
Time based collection
1. Select or enter the values for each option.
Option
Description
Pacing interval
Collect samples at regular time intervals, in hours and minutes (0:01–999:00).
Take first sample The program starts immediately or after the first interval.
Variable volume
Specify whether the sample volumes vary (flow based). If yes, refer to
on page 18. If no, refer to
Use variable volume
Sample pacing is constant time variable volume (CTVV).
1. Select or enter the values for each option.
Option
Description
Select unit
Select the flow units to use.
Map 4 mA input
Enter the flow rate that agrees with 4 mA input.
Map 20 mA input
Enter the flow rate that agrees with 20 mA input.
Average flow rate
Specify the average flow rate for the site (1–999,999). The units are specified by
the units selected above.
Time volume desired
Specify the total volume to be collected during the sample program (10–
10,000 ml).
18 English