Time based collection, Use variable volume, Figure 12 – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS Basic User Manual User Manual

Page 18

Advertising
background image

Option

Description

Advanced sampling

Goes to the advanced sampling menu. If necessary, configure the options in the
Advanced Sampling menu.

Run program

The creation of the basic sampling program finishes. The sampler prompts the user to
start or cancel the program.

Figure 12 Two samples per bottle (left) versus two bottles per sample (right)

Time based collection

1. Select or enter the values for each option.

Option

Description

Pacing interval

Collect samples at regular time intervals, in hours and minutes (0:01–999:00).

Take first sample The program starts immediately or after the first interval.

Variable volume

Specify whether the sample volumes vary (flow based). If yes, refer to

Use variable

volume

on page 18. If no, refer to

Do not use variable volume

on page 19.

Use variable volume
Sample pacing is constant time variable volume (CTVV).

1. Select or enter the values for each option.

Option

Description

Select unit

Select the flow units to use.

Map 4 mA input

Enter the flow rate that agrees with 4 mA input.

Map 20 mA input

Enter the flow rate that agrees with 20 mA input.

Average flow rate

Specify the average flow rate for the site (1–999,999). The units are specified by
the units selected above.

Time volume desired

Specify the total volume to be collected during the sample program (10–
10,000 ml).

18 English