Restore the default settings, Start or stop a program, Manual operation – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS Basic User Manual User Manual

Page 20: Volume calibration, Calibrate the liquid sensor

Advertising
background image

Option

Description

Time

The maximum time between samples (0:01–999:00). The timer starts again after each
sample is collected.

Sample volume

Enter the volume to be collected per sample. If bottles/sample mode is selected, each
bottle receives a full sample volume (100–10,000 mL). This option is skipped if variable
volume is selected.

Restore the default settings

The restore option sets all of the program parameters to the default settings.

1. Push MENU.
2. Go to PROGRAM SETUP>RESTORE.
3. Select YES.

Start or stop a program

Use the RUN/HALT PROGRAM key to start or stop a program. Up to 12 user-defined start/stop
times/dates can be set. A program must be stopped before the program or system setup can be
changed.

1. Push RUN/HALT PROGRAM.
2. Select one of the options.

Option

Description

Start

Starts the currently loaded program. The status changes to RUNNING.

Halt

Stops the program temporarily. The status changes to HALTED.

Resume

The program resumes from the point at which it was halted. The status changes to
RUNNING.

Start over

The program starts from the beginning. The status changes to RUNNING.

End program Stops the currently running program. The status changes to COMPLETE.

Manual operation

Use manual operation to collect a grab sample, move the distributor arm or operate the pump. For
additional information, refer to the expanded version of this manual.

Volume calibration

The sample volume can be calibrated automatically with the liquid sensor or manually. When the
liquid sensor is used for volume calibration, the sensor is adjusted to accurately measure all
programmed volumes. If the liquid sensor is disabled, the volume for samples, rinses and stormwater
must be calibrated manually.
For typical applications, tap water can be used for the volume calibration. If the sample composition
is significantly different from typical water samples, calibrate the sampler with the liquid to be
collected. All programs must be stopped before calibration.

Calibrate the liquid sensor

The accuracy of the liquid sensor varies with the type of liquid that is sampled. For example, the
volume of a highly turbid sample may not be as accurate as that of clear water. The liquid sensor can
be calibrated with the sample water to improve volume accuracy.

1. Put the intake tubing into the sample water.
2. Push MENU.

20 English

Advertising