Clean the instrument, Replace the desiccant, Fuse replacement – Hach-Lange SD900 Refrigerated Sampler, SD900 AWRS Basic User Manual User Manual

Page 23

Advertising
background image

Clean the instrument

C A U T I O N

Fire hazard. Do not use flammable agents to clean the instrument.

N O T I C E

Do not attempt to clean the controller compartment heater with liquids of any kind.

If water is not sufficient to clean the controller and the pump, disconnect the controller and move the
controller away from the sampler. Allow sufficient time for the controller and pump to dry before the
parts are re-installed and put back into service.
Clean the sampler as follows:

• Refrigerator—clean the condenser fins and coils as needed with a brush or vacuum.
• Sampler cabinet and tray—clean the internal and external surfaces of the sampler cabinet with a

damp cloth and mild detergent. Do not use abrasive cleaners or solvents.

Replace the desiccant

A desiccant cartridge in the controller absorbs moisture and prevents corrosion. Monitor the
desiccant color through the window (

Figure 13

). Fresh desiccant is yellow. When the color is green,

replace the desiccant.

1. Unscrew and remove the desiccant cartridge (

Figure 13

).

2. Remove the plug and discard the spent desiccant.
3. Fill the desiccant tube with fresh desiccant.
4. Install the plug.
5. Apply grease to the O-ring.
6. Install the desiccant tube in the controller.

Figure 13 Desiccant cartridge

1 Plug

3 O-ring

2 Desiccant tube

4 Desiccant window

Fuse replacement

Fuses are not user-serviceable items. The need for fuse replacement in controllers indicates severe
technical failure and is therefore considered to be a service activity. If a blown fuse is suspected,
contact Technical Support.

English 23