Getting started

33

Sunny icon

. The reagent calibration or manual entry of the

titre has been performed for all the reagents present in the
system. Everything is just right!

Cloudy icon

. The reagent calibration of one of the reagents in

the system should be performed within 12 or 24 hours.

Stormy icon

. The reagent calibration of one of the reagents

in the system has elapsed.

At least one of the reagents present in the system has not
been installed.

Check the sequence or method, (press 1 in the Main
window). The instrument prompts you to do the necessary
operations. You are guided step by step.

Question mark

. There is a problem in the editing of the

reagent system. You need to be in Supervisor mode to solve
the problem. Check the sequence/method
parameters of the reagent or the electrode.

Check the sequence or method, (press 1 in the Main
window). The instrument indicates the possible errors and
prompts you to correct them until ? disappears.