Chapter 3 – Hach-Lange TITRALAB 840_845 User Guide User Manual

Page 48

Advertising
background image

48

Chapter 3

Reference element and filling

Potential

Radiometer Analytical

solution

versus SHE

Electrodes

Hg/Hg

2

Cl

2

- Saturated KCl

+244 mV

pHC4000, pHC4001, pHC4006,

SCE: Saturated Calomel Electrode

XC601, REF401, REF421,
REF451, XR100, XR110,
XR130, XR150, MC408Pt

Hg/Hg

2

Cl

2

- 1M LiCl

+280 mV

REF921

NCE: Normal Calomel Electrode

Hg/Hg

2

SO

4

- Saturated K

2

SO

4

+651 mV

REF601, REF621, XR200,
XR230, MC602Pt, MC6091Ag

Hg/Hg

2

SO

4

- 1 M H

2

SO

4

+616 mV

Hg/HgO - 0.1 M KOH

+174 mV

XR400, XR430, XR440

Ag/AgCl - Saturated KCl

+199 mV

XR300, XR820, XC100, XC111,
XC120, XC161, XC200, XC250,

Ag/AgCl - 3 M KCl

+208 mV

pHC3001, pHC3005, pHC3006,
pHC3011, pHC3081, pHC3185,
REF321, REF361, MC3051Pt,
ISEC301F

Ag/AgCl - 1 M KCl

+235 mV

Ag/AgCl - 0.6 M KCl (sea water)

+250 mV

Red Rod - Saturated KCl

+199 mV

pHC2001, pHC2002, pHC2003,
pHC2005, pHC2011, pHC2015,
pHC2051, pHC2085, pHC2401,
pHC2441, pHC2501, pHC2601,
pHC2701, REF200, REF201,
REF251, REF261, MC2095Sb,

MC201Au-8

Advertising