Chapter 4
Run electrode calibration
Run electrode calibration
Run electrode calibration
Run electrode calibration
Run electrode calibration
Refer to Short-Form Reminder no. 4
.
1.
Select the method.
2.
Install the electrode group.
3.
Enter the Electrodes window.
4.
Select 1 Calibrate electrodes. Press
✓ to select the
electrode to be calibrated from the proposed list.
5.
Press 1 to start the calibration.
Run reagent calibration
Run reagent calibration
Run reagent calibration
Run reagent calibration
Run reagent calibration
Refer to Short-Form Reminder no. 5
.
1.
Select the method.
2.
Install the reagent to be calibrated.
3.
Install the electrode group to be used to perform the calibration.
4.
Enter the Reagents window.
5.
Select 1 Calibrate/Enter titre. Press
✓ and select the
reagent from the proposed list.
6.
Press 1 to start the calibration.
Run method
Run method
Run method
Run method
Run method
Refer to Short-Form Reminder no. 6.
1.
Select the method.
2.
Install the electrode then the reagent groups to be used to perform
the analysis.
3.
Display the Main window.
4.
Press 1 to start the analysis.