Run electrode calibration, Run reagent calibration, Run method – Hach-Lange TITRALAB 840_845 User Guide User Manual

Page 62

Advertising
background image

62

Chapter 4

Run electrode calibration

Run electrode calibration

Run electrode calibration

Run electrode calibration

Run electrode calibration

Refer to Short-Form Reminder no. 4

.

1.

Select the method.

2.

Install the electrode group.

3.

Enter the Electrodes window.

4.

Select 1 Calibrate electrodes. Press

✓ to select the

electrode to be calibrated from the proposed list.

5.

Press 1 to start the calibration.

Run reagent calibration

Run reagent calibration

Run reagent calibration

Run reagent calibration

Run reagent calibration

Refer to Short-Form Reminder no. 5

.

1.

Select the method.

2.

Install the reagent to be calibrated.

3.

Install the electrode group to be used to perform the calibration.

4.

Enter the Reagents window.

5.

Select 1 Calibrate/Enter titre. Press

✓ and select the

reagent from the proposed list.

6.

Press 1 to start the calibration.

Run method

Run method

Run method

Run method

Run method

Refer to Short-Form Reminder no. 6.

1.

Select the method.

2.

Install the electrode then the reagent groups to be used to perform
the analysis.

3.

Display the Main window.

4.

Press 1 to start the analysis.