Running analyses

61

4. Running analyses

4. Running analyses

4. Running analyses

4. Running analyses

4. Running analyses

✶ IMPORTANT ✶

Before pressing the Run key, check the following points listed below:

1.

Select the method by pressing 2 in the Main window.

2.

If Question marks "?" are present in the electrode and/or reagent
tabs
. Programming error due to missing electrode/reagent
parameters. Contact your supervisor.

3.

Connect/Check electrode(s).

Refer to Short-Form Reminder no. 1.

When installing/checking an electrode system, do not forget to
connect all the electrodes to the Titration Manager and disconnect
all the electrodes that are not used by the method.

4.

Install/Check reagent(s).

Refer to Short-Form Reminder no. 2

.

5.

Run the analysis when Sunny icons are visible in the electrode and
reagent tabs.

6.

Run an electrode calibration when a Stormy icon appears in the
electrode tab.

7.

Run a reagent calibration when a Stormy icon appears in the titrant
tab.

8.

If you are unable to display the "Run" command due to the presence
of Cloudy/Stormy icons in the Reagent and Electrode windows,
activate the "Check" command. The Titration Manager will
automatically guide you through the necessary operations required to
solve the problem(s).