Luminex IS Version 2.3 (IVD) User Manual
Page 24

Luminex IS Software Manual for Version 2.3- For In Vitro Diagnostic Use
x
MAP Technology
14
Figure 12 Acquisition Detail Toolbar
The Histogram, in the lower left section of the Acquisition Detail
tab, defaults to display the Doublet Discriminator (DD) on the X axis
and Events on the Y axis. Doublets appear when two microspheres
stick together, creating undesired results. When you enable the gate,
two vertical red, dashed lines appear to represent gate positions
determined by the template. You cannot change the gate position
during batch acquisition. You can change the gate positions before
data acquisition, after the system finishes acquiring data, or after
pausing the system. The change is visual. The data is still collected
according to the gate positions set in the assay template. The gate in
effect when the system collects data determines which values to use
to obtain the result. Applying a gate or changing a gate for existing
data does not change your calculated values. The gate positions, also
located in the lower corner of the histogram, are used while
collecting data are the numerical values selected in the template.