Managing assay lots – Luminex IS Version 2.3 (IVD) User Manual

Page 72

Advertising
background image

Luminex IS Software Manual for Version 2.3- For In Vitro Diagnostic Use

x

MAP Technology

62

Figure 34 Luminex Batch Setup for Dilution

4. To change the dilution factor, click the Dil. Factor field. The

white background indicates that you can enter text. Enter each
dilution factor as a decimal, not as a ratio.

5. Click Finish. The system saves the information you entered.

Managing Assay Lots

You can edit standard and control lot information. Once a lot is used,
changing or modifying it will prompt you for a new lot name. This
includes batches that have been set up, but not yet acquired.

You can modify known lot concentration values. If you change the
known concentration for a used lot, the system prompts you to enter
a new lot name. If you change the concentration values in an unused
lot, the system updates the lot with the new concentration values.

For assay reagents specified in templates, you can create new lots,
edit lot information, select pre-existing lots for reuse, import lots,
and export lots.

When editing lot numbers, follow these lot handling rules:

• If you have entered lot information for a template, but you

have not used the template to set up a batch, you can edit the
lot value.