tabs to observe the tests errors under the Errors tab and unknown
results under the Samples tab.
To view the next test in the batch click Next Test (F2). To view
the previous test, click Previous Test (F3). You may also click
on the test name in the left grid control.

Validating or

Invalidating

Standards and

Controls

You can invalidate or validate a standard or control in either of two
ways. In the Analysis window, use the associated buttons on the
bottom of the window, or right-click in the row containing the
standard or control you want to validate or invalidate.

Invalidating standards—You can invalidate or remove a standard if
doing so improves the curve fit. Observe caution when doing so.
Invalidating standards greatly affects the curve fit and subsequently
the sample results. For further instruction on assay standard curves
and the appropriateness of invalidating or removing standards,
contact the assay kit manufacturer.

Invalidating controls—You can invalidate or remove a control in
data analysis. However, Luminex does not recommend invalidating
controls. For further instruction on assay controls and guidelines
regarding accepting or rejecting control values, contact the assay kit
manufacturer.

To validate or invalidate a standard or control entry using the
right mouse-click procedure:

1. Right-click in the row containing the standard or control you

want to validate or invalidate. See Figure 48. Select the desired
menu item to apply. When invalidating, the Name box turns red
with an asterisk proceeding it. When validating, the Name box
reverts to black text.

1. Menu for multiple tests 2. Menu for single test

Figure 48 Invalidate and Validate Shortcut Menus

2. Click Recalc to recalculate the results if the Auto option was not

selected.