Multichannel Systems HiClamp Manual User Manual

Page 70

Advertising
background image

HiClamp Manual

64

6. Select the option Adjust Basket from the menu bar.

7. Select the basket position to Load Basket. The basket will move into the loading position and

the suction tube is lowered towards the basket. The suction tube will be used as the reference
point to which the basket needs to be aligned.

8. Use

the

two small Allen keys screws to bring the basket in the closest juxtaposition with

the suction tube but without contact.

9. Use

the

Backward or Forward button to align the basket to move the basket at the desired

position.

10. Press Done to close the CalibBasket window.

HiClamp is now ready for your next measurements.