2 percent weighing, Percent weighing – Nova-Tech PGW Series User Manual

Page 17

Advertising
background image

Page 17 of 44

9.4.2 Percent Weighing

Percent weighing will be done by defining a certain weight to be 100%. The
weight to be used can either be entered by the user or taken from a sample

• Press [Mode] and then the [Up] or [Down] key to select Percent

weighing, “PERCENT” will be displayed

• Press [Enter] to enter the function

• Display will show, “PERCENT SAMPLE”

• Press [Enter] to select the sample method

• When “LOAD 100 %“ is shown, add the sample

• Press [Enter] to set this weight to be 100%, the busy symbol will turn

on. When ready the display will show “100%”

• Remove the sample and place an unknown sample to display the

percentage weight

• To set another weight as 100%, press [Mode] and continue as before

• To manually enter a value to be set as 100%, press [Up] or [Down] key

when “PERCENT SAMPLE” is shown to select “PERCENT Ent Wt”

• Press [Enter] to select the manually entered weight method

• Enter the weight using the numeric entry method (see section 7.1)

• Place unknown sample to display the percentage weight

• To perform percent weighing with another sample press [Mode] and

continue as before

• To return to normal weighing, press [Esc]


Note: Percentage will be displayed to the maximum number of decimal
places based on the resolution of the balance. To increase or decrease by
one decimal place, press the [Up] or [Down] key respectively.

© Adam Equipment Company 2007

Advertising