• Next the beeper setting is displayed. Press [Up] or [Down] to scroll

through the options –

“BUZZER OFF” (Beeper set to off at all times)
“BUZZER In”

(Beeper will sound when the weight on the pan is
within the limit) or

“BUZZER Out” (Beeper will sound when the weight on the pan is

outside the set limits)

• Confirm the beeper setting by pressing [Enter]

• Press [Enter] again to start the Check Weighing

• When a weight is placed on the pan now, the display will indicate

whether the weight is below the LOW LIMIT between the LOW and
HIGH LIMITS or higher than the HIGH LIMIT using the capacity tracker,

• To perform check weighing with another sample press [Mode] and

continue as before

• To return to normal weighing, press [Enter]

9.4.4 Animal / Dynamic Weighing

The balance can be set to weigh animals or moving items, also known as
dynamic weighing. The balance will collect the weight over a period of time to
arrive at an average value and display the value till the operator resets the
balance. The actual weighing process can begin either automatically when
the weight is placed on the pan or when initiated by the operator. The
weighing unit to be used can be selected as normal using the [Unit] key,
before entering the Animal / Dynamic Weighing process.

Steps:

• Press [Mode] and then the [Up] or [Down] key to select Animal /

Dynamic Weighing. When “ANIMAL” is displayed press [Enter] to enter
Animal Weighing mode

• Press [Up] or [Down] to select “rUN” for starting the animal / dynamic

weighing using the method previously set or “SEtUP” to set up the
balance for animal weighing (see the section 9.4.4.1 Animal Weighing
Setup Parameters
)

© Adam Equipment Company 2007