12.0 ERROR CHECKING

During weighing the balance is constantly checking to see if the balance is
operating within the limited parameters. The errors likely to occur are:

A/D counts below lowest allowed value

A/D counts above highest allowed value

A/D not operating

Maximum

capacity

exceeded


Other errors may be detected during special functions or operations. These
will be described in the section that applies.

Error messages and the reasons are:

Concerning A/D counts

ERROR ADc UL

A/D counts below a limit

ERROR ADc OL

A/D counts above a preset limit

Concerning calibration

ERROR St

Calibration could not be completed
because the results were not stable

ERROR LO or ERROR HI

Calibration constant not within 20% of old
calibration constant

Concerning weighing

ERROR LO

Weight display is below zero by >4%max

ERROR HI

Weight is above maximum plus 90d

© Adam Equipment Company 2007