1 animal weighing setup parameters – Nova-Tech PGW Series User Manual

Page 21

Advertising
background image

Page 21 of 44

9.4.4.1

Animal Weighing Setup Parameters

• Press [Mode] and then [Up] or [Down] to select Animal Weighing.

When “ANIMAL” is displayed press [Enter] to enter the function

• Press the [Up] or [Down] key to select “rUN” or “SEtUP”

• Press [Enter] to select “SEtUP” to set up the balance for animal

weighing

• Use the [Up] or [Down] key to scroll through the options for setting up

the mode. The display will show “MODE AUTO” or “MODE MANUAL”.

• Select “MANUAL” or “AUTO”

Test time

• After [Enter] is pressed to select “MODE MANUAL” or “MODE AUTO”,

it will display “TEST TIME XX”, where XX is the number of seconds
over which the balance will average to compute the final weight as set
during last operation

• The XX can be changed from 1-200 using the numeric entry method

(see section 7.1)

• To confirm the desired value, press [Enter]

Delay

• Next, the display will show “DELAY XX” where XX is the time in

seconds taken by the balance before the sampling starts

• The XX can be changed from 0-200 using the numeric entry method

(see section 7.1)

• To confirm the desired value, press [Enter]

Threshold (During AUTO mode only)

• While in the AUTO mode, the display will next show “THRESHOLD XX

g” where XX is the minimum weight in gram of the item required by the
balance to start the process for animal weighing.

• The XX can be changed from 1 to 99 using the numeric entry method

(see section 7.1)

• To confirm the desired value, press [Enter]

© Adam Equipment Company 2007