NOTE: Percentage will be displayed to the maximum number of decimal
places based on the resolution of the balance. To increase or decrease by
one decimal place, press the

[Up] or [Down] key respectively.

9.4.3 Check Weighing

During weighing of a sample the balance can be set to show if the weight is
above or below an upper and a lower limit. The display will use the arrows
under the bar graph to show the check weighing is operating. The arrows
and bars between the arrows will indicate when the weight is below the lower
limit, between the limits or above the upper limit.


The buzzer can be set to be active when the weight is outside the limits
(below the lower or above the upper) or within the limits (above the lower and
below the upper limit), or turned off. If desired, only one limit needs to be set.
If only one limit is set the other limit is considered to be zero (lower) or the
maximum (upper).

The Check weighing is not active when the weight is less than 20d. This is
the minimum weight at which the indicator bars are displayed and the buzzer
sounds, if it is not set to OFF (see the next page).

Steps:

• Press [Mode] and then the [Up] or [Down] key to select Check

Weighing. When “CHECK” is displayed press the

[Enter] key to enter

Check Weighing mode

• Press [Up] or [Down] to set the “LOW LIMIT” to “On” or “OFF”

• Press [Enter] to proceed. If the “LOW LIMIT” was set to “On” the

display will show the current low limit in the unit, last used in the
weighing mode. You may change the limit using the numeric entry
method (see section 7.1).

• Press [Enter] to proceed

• If the “LOW LIMIT” was set to “OFF” or the setting of the low limit is

complete, then the display will change to “HIGH LIMIT”. Use

[Up] and

[Down] to set the “HIGH LIMIT” to “On” or “OFF”