Nova-Tech PW Series User Manual

Page 41

Advertising
background image

Page 41 of 48

© Adam Equipment Company 2007

Poor Linearity

Usually a mechanical
problem

Electronic Problems

Re-check repeatability

*Inspection of the flexures for damage
or loose hardware may be required

*Use the Linearity Function in the
service menu to reset linearity

*A problem in the analogue circuit board
or power supplies can cause poor
linearity. Make sure all mechanical
problems have been eliminated first

OTHER PROBLEMS

Cannot calibrate

Zero shifted more
than allowed

Calibration timeout

*Check all flexures for damage

*Reset factory calibration

*Verify linearity and repeatability

*The balance may be unstable. Verify
stability as above. Try using a more
aggressive filter

Calibration weight
motor does not stop

*Check the cables to the motor, try
plugging the balance into the power
again

*Look for friction in the calibration
weight movement

*Check the opto-coupler that controls
the motor position.

RS-232 not working

Doesn’t print

Check parameters match the device
connected

Verify cable is correct

*RS-232 circuits damaged

Display dark, keys
beep

Display contrast poor

Cable unplugged or
damaged

*Check the cables to the display

*Replace the display which could be
damaged

*To be carried out by authorised technicians only.