CAL REPORT

On = Enabled. Prints out Calibration report after
successful calibration
OFF = Disabled

TIME CAL

On = Enabled. Select time from 1 to 24 hours.
Default setting is 4 hours
OFF = Disabled

TEMP CAL

On = Enabled. If set to On, set the temperature
variation from 0.2 to 4°C
OFF = Disabled

INT CAL

YES = Use Internal calibration mass as displayed or
adjust for accurate Internal Calibration (see section
13.5.1)
NO = Use external calibration mass


• Press [Mode] to advance to setting of the next menu which is

“PASSCODES” or

[Esc] to return to normal weighing.



13.5.1

INTERNAL CALIBRATION MASS ADJUSTMENT

Balance model

Internal Calibration mass (approx.)

PW 124

100g

PW 184

100g

PW 214

100g

PW 254

100g

NOTE: The value of the internal mass is not an exact number, such as 100g.
It must be set by comparing with an external mass. The exact value can be
set, provided this feature is enabled at the factory.

PROCEDURE

• Make sure the balance is level and has been ON for 4 hours or more at

a stable temperature.

• Calibrate the balance twice. This helps to reduce any stresses in the

mechanics.

• To verify the accuracy of this internal calibration, place an external

mass approximately equal to the same internal calibration mass.
Record the result.