Operating Instructions
Model 9110TH NOx Analyzer
Teledyne Analytical Instruments
85
In order to compensate for CPU clocks which run fast or slow, there is a variable to
speed up or slow down the clock by a fixed amount every day.
To change this variable, press:
SAMPLE
ENTER SETUP PASS : 818
8 1 8
ENTR
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SETUP X.X 0 ) DAS_HOLD_OFF=15.0 Minutes
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
3x EXIT returns
to the main SAMPLE display
Enter sign and number of seconds per
day the clock gains (-) or loses (+).
SETUPX.X 1 ) MEASURE_MODE=NOX-NO
PREV
NEXT
JUMP
EDIT PRNT EXIT
SETUP X.X
7) CLOCK_ADJ=0 Sec/Day
PREV JUMP
EDIT
PRNT EXIT
SETUP X.X
CLOCK_ADJ:
0 Sec/Day
+ 0 0
ENTR
EXIT
SETUP X.X
7) CLOCK_ADJ=0 Sec/Day
PREV NEXT JUMP
EDIT PRNT
EXIT
Continue to press NEXT until …