Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 144

Advertising
background image

Model 9110TH NOx Analyzer

Operating

Instructions

Teledyne Analytical Instruments

124

The available gas concentration DAS parameters for output via the 9110TH/M analog
output channels are:

Table 4-23: Analog Output DAS Parameters Related to Gas Concentration Data

REPORTING

RANGE

PARAMETER

NAME

1

DESCRIPTION

NO

x

Range 1

(LOW)

NXCNC1

Concentration

NXSLP1

Slope

NXOFS1

Offset

NXZSC1

Concentration during calibration, prior to computing new slope and offset

NO

x

RANGE 2

(HIGH)

NXCNC2

Concentration

NXSLP2

Slope

NXOFS2

Offset

NXZSC2

Concentration during calibration, prior to computing new slope and offset

NO Range 1

(LOW)

NOCNC1

Concentration

NOSLP1

Slope

NOOFS1

Offset

NOZSC1

Concentration during calibration, prior to computing new slope and offset

NO RANGE 2

(HIGH)

NOCNC2

Concentration

NOSLP2

Slope

NOOFS2

Offset

NOZSC2

Concentration during calibration, prior to computing new slope and offset

NO

2

Range 1

2

(LOW)

N2CNC1

Concentration - Computed with data from NO

x

Range 1 and NO Range 1

N2ZSC1

Concentration during calibration, prior to computing new slope and offset

NO

2

RANGE 2

2

(HIGH)

N2CNC2

Concentration - Computed with data from NO

x

Range 2 and NO Range 2

N2ZSC2

Concentration during calibration, prior to computing new slope and offset

O

2

Range

3

O2CONC

3

Concentration

O2OFST

3

Slope

O2SLPE

3

Offset

O2ZSCN

3

Concentration during calibration, prior to computing new slope and offset

1

Parameters are not listed in the order they appear on the DAS list (see Table A-6 or Appendix A.5 for the proper order of the full list of

parameters)

2

Since NO

2

values are computed rather than measured directly, no separate slope or offset exist.

3

Only available on instruments with O

2

sensor options installed.