Electronic operation, Pmt preamp pca, Pc 104 cpu card – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual

Page 299: Mother board

Advertising
background image

Troubleshooting & Repair

Model 9110TH NOx Analyzer


Teledyne Analytical Instruments

279

8.5. ELECTRONIC OPERATION

Figure 8-15 shows a block diagram of the major electronic components of the
9110TH/M.

Pneumatic

Sensor

Board

Sample

Pressure

Sensor

Vacuum

Pressure

Sensor

O

3

Flow Sensor

Analog Outputs

Status Outputs:

1 – 8

Control Inputs:

1 – 6

PC 104

CPU Card

Disk On

Module

Flash Chip

C OM1 (RS–232 ONLY)

COM2 (RS–232 or RS–485)

Power-Up

Circuit

I

2

C Bus

Analog

Sensor Inputs

Box

Temp

Thermistor

Interface

REACTION CELL

TEMPERATURE

MOLYBDENUM CONVERTER

TEMPERATURE

PMT

Temperature

Sensor

A1

A2

A3

Optional

4-20 mA

MOTHER

BOARD

A/D

Converter(

V/F)

PC 104

Bus

External

Digital I/O)

Analog

Outputs

(D/A)


RELAY

BOARD

I

2

C Status

LED

PUMP

(Externally Powered)

A4

CPU Status

LED

NO/NO

x

Valve

Autozero

Valve

Sample Cal

Valve Option

Option

Reaction Cell

Heater

Molybdenum

Converter Heater

PMT TEC

PMT

MOLYBDENUM CONVERT ER

TEMPERATUR E SIGNAL

TEC Drive

PCA

Internal

Digital I/O

EL

EC

T

R

IC

T

E

ST

C

O

N

T

R

O

L

O

P

TI

C

TE

S

T

CO

N

T

R

O

L

P

M

T O

U

TP

U

T

(

PM

T

D

E

T

)

O

2

Sensor

Option

H

IG

H

V

O

LT

AG

E

PO

W

E

R

S

U

PP

L

Y

LE

V

E

L

PM

T

T

E

M

PER

A

T

U

R

E

PMT

PREAMP PCA

O

2

OPTION

TEMPERATURE

Display

Touchscreen

LVDS

transmitter board

Analog RS232 COM2 USB Ethernet
IN

Male

Female COM port

USB

or USB

(I

2

C Bu s)

Figure 8-15:

9110TH/M Electronic Block Diagram

The core of the analyzer is a microcomputer (CPU) that controls various internal
processes, interprets data, calculates data, and reports results using specialized firmware
developed by Teledyne. It communicates with the user, receives data from and issues

Advertising