Instrument id code – Teledyne 9110TH - Nitrogen Oxides Analyzer User Manual
Page 184

Model 9110TH NOx Analyzer
Operating
Instructions
Teledyne Analytical Instruments
164
To assign or reset the status flag bit assignments, press:
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS DIAG ALRM
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK
MORE
EXIT
SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
< TST TST > CAL
SETUP
SETUP X.
HESSEN STATUS FLAGS
<SET
SET>
EDIT
EXIT
SETUP X.X
COMMUNICATIONS MENU
ID
HESN
COM1
COM2
EXIT
SETUP X.
PMT DET WARNING: 0002
PREV NEXT
EDIT
PRNT
EXIT
Repeat pressing SET> until …
SETUP X.
SYSTEM RESET: 0000
PREV NEXT
EDIT
PRNT
EXIT
<CH and CH>
move the [ ]
cursor left and
right along the
bit string.
Repeat pressing NEXT or PREV until the desired
message flag is displayed. See Table 6-29.
For example …
ENTR key accepts the
new settings
EXIT key ignores the new
settings
SETUP X.
SYSTEM RESET: [0]000
<CH CH>
[0] ENTR EXIT
Press the [?] key repeatedly to cycle through the available character set: 0-9
Note: Values of A-F can also be set but are meaningless.
4.15.4.7. Instrument ID Code
Each instrument on a Hessen Protocol network must have a unique ID code. The
9110TH/M has a default ID of either 0 or 200. To change this code see Section 4.11.1