4.11.3. COMMUNICATION MODES, BAUD RATE AND PORT TESTING
Use the SETUP>MORE>COMM menu to configure COM1 (labeled RS232 on 
instrument rear panel) and/or COM2 (labeled COM2 on instrument rear panel) for 
communication modes, baud rate and/or port testing for correct connection. 
4.11.3.1. COM Port Communication Modes
Each of the analyzer’s serial ports can be configured to operate in a number of different 
modes, which are listed in the following table. Each COM port needs to be configured 
independently. 
Table 4-11: COM Port Communication modes
MODE
1
ID
DESCRIPTION
QUIET
1
Quiet mode suppresses any feedback from the analyzer (DAS reports, and warning 
messages) to the remote device and is typically used when the port is communicating 
with a computer program such as APICOM. Such feedback is still available but a 
command must be issued to receive them. 
COMPUTER
2
Computer mode inhibits echoing of typed characters and is used when the port is 
communicating with a computer program, such as APICOM. 
SECURITY
4
When enabled, the serial port requires a password before it will respond. The only 
command that is active is the help screen (? CR). 
HESSEN
PROTOCOL
16
The Hessen communications protocol is used in some European countries. TAI part 
number 02252 contains more information on this protocol. 
E, 7, 1
2048
When turned on this mode switches the com port settings 
 from 
No parity; 8 data bits; 1 stop bit
to
Even parity; 7 data bits; 1 stop bit
RS-485
1024
Configures the COM2 Port for RS-485 communication. RS-485 mode has precedence 
over multidrop mode if both are enabled. When the COM2 port is configured for RS-485 
communication, the rear panel USB port is disabled. 
MULTIDROP
PROTOCOL
32
Multidrop protocol allows a multi-instrument configuration on a single communications 
channel. Multidrop is an option requiring a special PCA and the use of instrument IDs. 
ENABLE
MODEM
64
Enables sending a modem initialization string at power-up. Asserts certain lines in the 
RS-232 port to enable the modem to communicate. 
ERROR
CHECKING
2
128
Fixes certain types of parity errors at certain Hessen protocol installations.
XON/XOFF
HANDSHAKE
2
256
Disables XON/XOFF data flow control also known as software handshaking.
HARDWARE
HANDSHAKE
8
Enables CTS/RTS style hardwired transmission handshaking. This style of data 
transmission handshaking is commonly used with modems or terminal emulation 
protocols as well as by Teledyne Instrument’s APICOM software. 
HARDWARE
FIFO
2
512
Improves data transfer rate when on of the com ports.
COMMAND
PROMPT
4096
Enables a command prompt when in terminal mode.
1
Modes are listed in the order in which they appear in the
SETUP
 MORE  com  COM[1 OR 2]  MODE menu
2
The default sting for this feature is ON. Do not disable unless instructed to by TAI Technical Support personnel.