Operating Instructions

Model 9110TH NOx Analyzer

Teledyne Analytical Instruments

109

4.12.1. SETTING THE GAS MEASUREMENT MODE

In its standard operating mode the 9110TH/M measures NO, NO

2

and NO

x

. It can also

be set to measure only NO or only NOX. To select one of these three measurement
modes, press:

SETUP X.X

1 ) MEASURE_MODE=NOX-NO

PREV

NEXT

JUMP

EDIT

PRNT EXIT

EXIT ignores the new

setting.

ENTR accepts the

new setting.

SAMPLE

ENTER SETUP PASS : 818

8 1 8

ENTR

EXIT

SAMPLE A1:NXCNC1=100PPM NOX=XXX.X

< TST TST > CAL

SETUP

SETUP X.X

MEASURE MODE: NOX-NO

PREV

ENTR EXIT

SETUP X.X

MEASURE MODE: NOX

PREV

NEXT

ENTR EXIT

SETUP X.X

MEASURE MODE: NO

NEXT

ENTR EXIT

NOX-NO mode is the

default mode for the

200EH/M

Press the PREV

and NEXT buttons

to move back and
forth between gas

modes

SETUP X.X

PRIMARY SETUP MENU

CFG DAS

RNGE

PASS CLK

MORE

EXIT

SETUP X.X

SECONDARY SETUP MENU

COMM

VARS

DIAG

EXIT

SETUP X.X

0 ) DAS_HOLD_OFF=15 minutes

NEXT

JUMP

EDIT

PRNT EXIT