Operating Instructions
Model 9110TH NOx Analyzer
Teledyne Analytical Instruments
101
5. In the instrument’s SETUP>MORE>COMM>COM2 menu, make the following settings:
Baud Rate: 115200 
COM2 Mode Settings: 
Quiet Mode 
ON
Computer Mode
ON
MODBUS RTU
OFF
MODBUS ASCII
OFF
E,8,1 MODE
OFF
E,7,1 MODE
OFF
RS-485 MODE
OFF
SECURITY MODE
OFF
MULTIDROP MODE
OFF
ENABLE MODEM
OFF
ERROR CHECKING
ON
XON/XOFF HANDSHAKE OFF 
HARDWARE HANDSHAKE OFF 
HARDWARE FIFO 
ON
COMMAND PROMPT
OFF
6. Next, configure your communications software, such as APIcom. Use the COM port
determined in Step 4 and the baud rate set in Step 5. The figures below show how 
these parameters would be configured in the Instrument Properties window in 
APIcom when configuring a new instrument. See the APIcom manual (PN 07463) 
for more details. 
Note
USB configuration requires that instrument and PC baud rates match; check the 
PC baud rate and change if needed. Using the USB port disallows use of the rear 
panel COM2 port except for multidrop communication.